ELISA Kits COVID-19 IgG
Severe acute respiratory syndrome Coronavirus 2 (SARS-CoV-2) is the name given to the new coronavirus that originated in 2019 and is the causative agent of COVID-19 (the disease caused by the SARS-CoV-2 virus). The virus was originally known as the 2019-nCoV nomenclature. SARS-CoV-2 is a new coronavirus strain that has not been previously identified in humans. The genome of these viruses consists of single-stranded RNA protected by two protein coatings - nucleocapsid and capsid. Its genetic characteristics are significantly different from SARS-CoV and MERS-CoV.
EDI™ Novel Coronavirus COVID-19 IgG ELISA Kit
is intended for qualitative measurement of the human anti-COVID-19 IgG antibody in serum.
This assay utilizes t
he method of
microplate based
indirect
enzyme-linked immunosorbent assay
.
PRINCIPLE OF THE TEST
Assay controls and diluted serum samples are added to the microtiter wells of a microplate that was coated with COVID-19 recombinant full length nucleocapsid protein. After the first incubation period, the unbound protein matrix is removed with a subsequent washing step. A horseradish peroxidase (HRP) labeled polyclonal goat anti-human IgG tracer antibody is added to each well. After an incubation period, an immunocomplex of "COVID-19 recombinant antigen – human anti-COVID-19 IgG antibody - HRP labeled anti human IgG tracer antibody" is formed if there is specific coronavirus IgG antibody present in the tested specimen. The unbound tracer antibody is removed by the subsequent washing step. HRP tracer antibody bound to the well is then incubated with a substrate solution in a timed reaction and then measured in a spectrophotometric microplate reader. The enzymatic activity of the tracer antibody bound to the anti-COVID-19 IgG on the wall of the microtiter well is proportional to the amount of the anti-COVID-19 IgG antibody level in the tested specimen.
MATERIALS AND REAGENTS
COVID-19 antigen coated microtiter plate
COVID-19 IgG Sample Diluent
HRP labeled Anti-human IgG Tracer Antibody
ELISA Wash Concentrate
ELISA HRP Substrate
ELISA Stop Solution
COVID-19 IgG Negative Control
COVID-19 IgG Positive Control
PERFORMANCE CHARACTERISTICS
Limit of Detection
No international standardized units are available for COVID-19. A positive sample was serially diluted and the limit of detection was determined to be not higher than 5 U/mL.
Repeatability
The assay control is tested in 10 replicates with a CV of OD values less than 15%.
Reproducibility
Three lots were tested with the same samples 10 times with a CV less than 20%.
Cross-Reactivity
Panels were studied with a minimum of five confirmed disease state samples. No interference was observed for the following disease or infectious agents:
Anti-influenza A
Anti-influenza B
Hepatitis C (HCV)
Antinuclear Antibodies (ANA)
Respiratory Syncytial Virus (RSV)
Class Specificity
This assay does not show any cross reaction to IgМ.
SENSITIVITY AND SPECIFITY
The diagnostic sensitivity is 100%.
The diagnostic specificity is 100%.
The negative predictive value is 100%.
The positive predictive value is 100%.